HJT-80KV insulation oil dielectric strength tester is a fully automatic transformer oil BDV test kit designed with advanced technologies to measure the dielectric strength of transformer oil
HJT-80KV Insulation Oil Breakdown Voltage Analyzer is a fully automatic transformer oil BDV test kit designed with advanced technologies to measure the dielectric strength of transformer oil. Our transformer oil dielectric strength tester is available at reasonable price with option to choose from 60/80/100kV range.
Main functions and features
1.The Oil BDV tester is controlled by a large-capacity microcontroller and works stably and reliably;
2.The Oil BDV tester is equipped with a wide range watchdog circuit to prevent crashes;
3.Multiple operation options. The instrument program is equipped with IEC156 standard, ASTM D877
4.The instrument oil cup is cast in one step using special glass, which eliminates interference phenomena such as oil leakage;
5.The unique high-voltage side sampling design of the instrument allows the test value to directly enter the A/D converter,
avoiding errors caused in the analog circuit and making the measurement results more accurate;
6.The instrument has over-current, over-voltage, short-circuit and other protection functions inside, and has strong
anti-interference ability and good electromagnetic compatibility;
7.Portable structure, easy to move, easy to use indoors and outdoors.
Customized support |
OEM |
Place of Origin |
China |
Brand Name |
HOPU |
Model Number |
HJT-80KV |
Power |
200W |
Product name |
insulating oil dielectric strength tester |
Output voltage |
0~80KV(100kv customizable) |
Voltage distortion rate |
<3% |
Voltage raising speed |
0.5~5 KV/S (adjustable) |
Standing time |
15min (adjustable) |
Boost interval |
5 min (adjustable) |
Boosting frequency |
1~9 time(s) (optional) |
Booster capacity |
1.5KVA |
Measuring accuracy |
±3% |
Supply voltage |
AC220V±10% 50Hz±1 Hz |
Applicable temperature |
0℃~45℃ |
Applicable humidity |
<75%RH |
Overall dimension |
465×385×425 MM |